當前位置:上海境宣實業(yè)有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 190U二丙酮醇氣體檢測管190U DIACETONE ALCOHOL 10-250ppm
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
二丙酮醇氣體檢測管190U DIACETONE ALCOHOL 10-250ppm
1. PERFORMANCE
1) Measuring range :10-250 ppm
Number of pump strokes 3(300ml)
2) Sampling time :4.5 minutes/3 pump strokes
3) Detectable limit :1 ppm
4) Shelf life :2 years
5) Operating temperature :10~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Graduations printed on the tube are calibrated by Ethyl cellosolve at 3 pump strokes
and Diacetone alcohol concentration is determined by using a conversion chart.
8) Colour change :Yellow→Pale blue
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
(CH3)2C(OH)CH2COCH3+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
200
150
500
400
300
100
50
20
5
300ml
ppm
(White)
DEHUMIDIFIER
Tube No.
DIACETONE ALCOHOL
92
Substance Interference Coexistence
Alcohols FIG.1 Similar stain is produced. Higher readings are given.
Halogenated hydrocarbons 〃
Aliphatic hydrocarbons Whole reagent is discoloured to It the top of Pale blue stain is clear,the
Pale brown. accuracy of readings is not affected.
Aromatic hydrocarbons 〃〃
Esters 〃〃
0
20
30
40
50
10
Ethanol〔ppm〕?
FIG.1 Influence of Ethanol
Diacetone indicating value 〔ppm〕
50 100
Coexistence
Interference
Diacetone alcohol(ppm)?
90U Tube reading(ppm)
請輸入賬號
請輸入密碼
以上信息由企業(yè)自行提供,信息內容的真實性、準確性和合法性由相關企業(yè)負責,化工儀器網對此不承擔任何保證責任。
溫馨提示:為規(guī)避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。